Skip to main content

What is the FRAME?

The FRAME (Fast Rigid Automated Microscope Engine) is an inverted, motorised microscope built from the openUC2 cube system. The optics stay fixed; the sample moves.

FRAME ID10–ID20

For people deciding whether a FRAME fits their work.

The core idea: fixed optics, moving sample​

The objective does not move for focus. The sample plate moves in X, Y and Z, and a slider changes between two objectives. Everything below the objective stays in one position relative to the objective's back focal plane.

That matters for anything aligned to that plane: structured illumination, a confocal or laser-scanning add-on, a focus lock. Such a module can be aligned once on the bench, slid in, and stays aligned through a z-stack.

Modular by design​

  • Below the stage sits an optics drawer of 3 x 3 openUC2 cubes in one or two layers: fold mirror, tube lens, filter and dichroic cubes, camera. Cubes can be exchanged or rearranged.
  • Between the drawer and the stage, a 33 mm slot takes flat slide-in modules such as the focus lock.
  • One 12 V supply and one CAN bus feed every motor, lamp and laser board. New modules plug into the same bus.

Open source​

Hardware, firmware (UC2-ESP32) and software (ImSwitch) are open. You can change the optical path, add a board to the bus, or control the instrument from Python or the REST API.

Who it is for, and its limits​

  • Automated imaging of slides, well plates and dishes: tile scans, time-lapse, multi-position experiments.
  • Brightfield, darkfield and DPC with the LED rings; fluorescence with fibre-coupled lasers or LEDs.
  • Research use only. It is not a medical or in-vitro diagnostic device.

Limits of units ID10–ID20 to know before you buy or plan an experiment:

LimitValue
Field of viewabout 10 mm image circle; 0.74 x 0.49 mm at 10x with the standard camera
Stage repeatability without encoderwithin about ±7 µm over a day of moves
Focus travel8 mm
Temperatureprinted parts limit use to 38 °C

Details: Hardware ID10–ID20.