ImSwitch will not shut down the machine
- Symptom / cause / fix. Cross-link: shutdown tutorial.
Misleading status / error message
- Symptom / cause / fix.
GUI switching bug
- Symptom / cause / fix.
Live view is frozen
- Symptom: the live image stops updating while the stage and the lamp still react.
- Cause: the video stream to the browser stalled.
- Fix: stop and start the stream in the Live View. If it stays frozen, reload the page.
A scan pauses for several seconds
- Symptom: a scan stops between tiles and continues by itself.
- Cause: not confirmed. Seen at a commissioning in November 2026 while frames were written to the micro SD card.
- Fix: the data is complete. Write to a USB drive or an SSD if it happens often, and report it with the log.