Optical layout of the FLIM module (OPM-0004)

Information-oriented. Numbers from the CAD assembly OPM - 0004 - FLIM 488 and from the Optiland model of it (simulation).

Elements​

The scan layer is the upper layer of the module (cube centres at z = 55 mm in the CAD). Cells are given as (row, column) in the drawing above: rows top to bottom, columns left to right.

elementCAD modulecellkey data
excitation fibre collimatorASS-2084 CUBECCADJFIBHOL+SHLFILLCKVER(2, 2)FC/PC collimator f = 11.2 mm on an XY (eccentric) holder; excitation filter slider SUB-0093
junction dichroicASS-2078 CUBKINDICSPL45°(2, 1)kinematic 45° mount, tip / tilt; reflects the 488 nm excitation into the galvo, transmits the fluorescence
detection fibreASS-2083 CUBECCFIBHOL+SHLFILLCKVER(1, 1)coupling lens + fibre on an XY holder; emission filter slider SUB-0087
galvo pairASS-2075 CUBGLV(3, 1)mirrors 10.2 × 6 mm (CAD); beam in along x on mirror A (7.4 mm above the cube centre), down to mirror B (cube centre), out along the row
scan lensASS-2073 CUBLEND25F50(3, 2)f50, Ø25.4 plano-convex (R 25.84 mm, 4.69 mm, N-BK7), convex side toward the galvo
empty cubesASS-2000(3, 3)-(3, 5)spacers
tube lensASS-2074 CUBLEND25F160(3, 6)f160, Ø25.4 plano-convex (R 82.62 mm, 2.95 mm, N-BK7), insert offset +9.5 mm, convex side toward the objective
splitterASS-2077 CUBKINSPL45°(3, 7)25 × 25 × 1 mm plate in a kinematic 45° mount; folds the beam up through the top-plate port to the objective
top-plate portPRT-1052 + PRT-4007above (3, 7)M37 port with a glass disc, clear Ø34
What is in your module?

The CAD lists plano-convex f50 and f160 lenses. Some modules carry an f50 achromat and an f180 or f200 tube lens instead. That changes the magnification (3.6x or 4x), the beam size in the pupil and the field per degree. The simulation covers both; check the engraving on your lenses.

Distances along the beam​

Unfolded, measured from mirror B (the plane the telescope images into the pupil):

planedistance
mirror A-7.4 mm (before mirror B)
mirror B (scan pivot of z)0
scan lens (curved vertex)49.5 mm
intermediate image100.6 mm
tube lens (first vertex)256.8 mm
splitterabout 299 mm (39 mm after the tube lens)
top-plate portabout 330 mm
FRAME stage plate opening (about Ø15.5)about 372 mm
objective changer opening (Ø14)about 384 mm
image of mirror B = objective back focal plane418 mm (158 mm after the tube lens, inside the objective)

The positions of the FRAME openings are taken from the FRAME CAD and are accurate to a few millimetres.

Beam and scan​

quantityvalue
wavelength488 nm
beam diameter at the galvo (1/e²)4 λ f / (π MFD) = 2.0 mm for f 11.2 mm and a 3.5 µm mode-field diameter
relay magnificationf_TL / f_SL = 3.2
beam diameter in the objective pupil6.4 mm (2 mm beam), 12.8 mm (4 mm beam)
optical angle2 × mechanical mirror angle
angle in the pupiloptical angle / 3.2
spot shift at the sample245 µm per optical degree (f_obj 45 mm, 4x), 110 µm (20 mm, 10x), 55 µm (10 mm, 20x)
image of mirror A7.4 × 3.2² = 76 mm beyond the BFP: the x scan walks in the pupil by about 0.4 mm per optical degree

Apertures that can clip the beam​

aperturesize seen by the beamclips first when
galvo mirror A, across its shaft6 mm × cos 45° = 4.2 mm (8 × 5 mm mirrors: 3.5 mm)the beam is wider than about 3 mm, or off-centre
galvo mirror B, across its shaftsamesame
scan lens mountØ about 23 mmoptical angles beyond about ±9°
tube lens mountØ about 23 mmz angles beyond about ±8° (large beams earlier)
splitter 25 × 25 mm at 45°25 mm × 17.7 mm (the long side across the fold)z angles beyond about ±8°
FRAME stage plate openingabout Ø15.5beams wider than 3 mm at the galvo, and beams off-centre
objective changer openingØ14same
objective pupil2 NA f_obj: 9 mm (4x / 0.10), 12 mm (10x / 0.30), 8 mm (20x / 0.40)x angles beyond about ±5° (walk), or when the beam is decentred

Adjustable elements​

elementadjustsused in
excitation fibre holder XYposition of the excitation beamtutorial step 3 (near)
detection fibre holder XYposition of the detection axis; peak-up of the countstutorial steps 3, 4, 11
junction dichroic tip / tiltdirection of the excitation into the galvotutorial step 3 (far); after transport
galvo centre (DAC)scan centre = optical axistutorial step 5
splitter tip / tiltdirection into the objective: field centre and pupil centringtutorial step 8; after transport