How to re-align the scanner after transport
Task-oriented. Use this when the system worked before and now gives a weak, cut or off-centre beam.
The lenses and the galvo sit in cubes on the 50 mm grid and do not drift. After transport you touch two kinematic cubes and one software setting:
| what | moves | how you see it |
|---|---|---|
| junction dichroic (next to the excitation fibre) | the excitation beam's direction into the galvo, so its position on the galvo mirrors and in the objective pupil | spot position on the mirrors; overlap with the red detection laser far away |
| splitter under the objective (end of the scan row) | the beam's direction into the objective, so the field centre and the beam position in the pupil | spot on a card at the objective position |
| galvo centre (ImSwitch, DAC X / Y) | the scan centre | position of the scan field in the camera image |
Before you start
- Goggles on, laser at minimum power, objective removed.
- Galvo powered and parked at your centre values (from the tutorial, step 5; default 2048 / 2048).
Step 1 - Beam on the galvo mirror centres
Look onto the two galvo mirrors from above through the open galvo cube.
- Centred on both: go to step 2.
- Off-centre, especially across the short (6 mm) side: correct it with the junction dichroic. Then check that the excitation still overlaps the red detection laser far away (tutorial step 3, the "far" part). If not, re-do that step.
Step 2 - Beam at the objective position
Put the module into the FRAME (objective still out) and place a card on the objective changer.
- Round and centred on the 14 mm opening: go to step 3.
- Off-centre: centre it with the splitter's tip / tilt screws, in small steps.
- D-shaped, crescent or missing: work through Fix a clipped or off-centre beam first.
Step 3 - Scan field in the image
Objective in, fluorescent slide in focus, widefield camera on.
- Send the beam to the centre and look where the spot is in the camera image.
- Small offsets (up to about 1° optical, which is about 250 µm with the 4x): use the galvo centre in software.
- Larger offsets: the splitter is tilted. Correct it there (step 2), not with the galvo: a large galvo offset pushes the beam towards the edges of the lenses and the pupil.
What each adjustment costs
From the simulation, with the 4x and a 2 mm beam:
- A 0.5° tilt of the splitter turns the beam by 1°. That moves the spot about 800 µm in the sample and moves the beam 1.8 mm across the objective pupil.
- A 2° optical offset of the galvo centre costs little at the centre (8 %), but about two thirds of the peak at one edge of the field.
Details: Optical quality and limits.
Step 4 - Peak up the detection
With a bright uniform sample, scan a small area and maximise the mean count rate with the detection fibre holder's XY (tutorial step 11).
Don't
- Don't move the scan lens, the tube lens or the galvo cube. They are on the grid and are not the cause.
- Don't adjust both kinematic cubes at once: change one, check, then the next.
- Don't use the galvo offset to compensate a beam that sits on a mirror edge.